UV-Spectrophotometry and RP-HPLC methods for the simultaneous estimation of acetaminophen: Validation, comparison and application for marketed tablet analysis in South West, Nigeria
ABSTRACT:
In the present study UV-Spectrophotometry
and RP-HPLC methods were validated for the simultaneous analysis of
acetaminophen in marketed tablets. The methods were validated in terms of
linearity, sensitivity (Detection limit and Quantification limit) accuracy (%
Recovery), precision (inter day, intraday and reproducibility). Both the
methods were linear (r2 = 0.9993 for UV method and 0.9995 for HPLC) and
accurate (% recovery was 99.48 % - 101.42 % for UV method and 101.85 % - 102.35
% for HPLC method). The detection limit and quantification limit were 0.192
µg/ml and 0.640 µg/ml for UV method and 0.0155 µg/ml and 0.0518 µg/ml. The
method was also found precise (% RSD < 5%) and robust. Assay of five
marketed brands of paracetamol were determined by both the methods and no
statistically significant difference was noticed between the assay obtained
from UV-Spectrophotometry and RPHPLC methods by paired t - test at 5 %
significance level. The results obtained from the mean percentage analysis of
paracetamol tablets (%) containing 500 mg of acetaminophen shows that the mean
percentage determined in three replicate analyses is more than the claimed
amount by the manufacturers. The two methods were found to be linear,
quantitative, reproducible and could be used as a more convenient, efficient and
economical method for the trace analysis of drug in raw material, tablets and
in biological fluids.